      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74567",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide (including Puerto Rico) and the countries of: Australia, Canada, Colombia, Germany, Guatemala, Netherlands, Singapore, Tunisia, , Spain",
      "recall_number": "Z-2274-2016",
      "product_description": "Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA.    Product Usage:  For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.",
      "product_quantity": "62 units total (33 units in US)",
      "reason_for_recall": "A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.",
      "recall_initiation_date": "20160205",
      "center_classification_date": "20160722",
      "termination_date": "20170117",
      "report_date": "20160803",
      "code_info": "AFB13H005, AFB13H008, AFB13H012, AFB13K002, AFB13K005, AFB13K008, AFB13J002, AFB13J003, AFB13J004, AFB13J005, AFB13J007, AFB15J001, AFB15C010, AFB13K009, AFB14H001, AFB15C008, AFB15C003, AFB15C005, AFB15C004, AFB15C006, AFB14H005, AFB14H003, AFB13K004, AFB13K006, AFB13K007, AFB13J008, AFB13K001, AFB13J006, AFB13J010, AFB13H013, AFB13H009, AFB13H003, AFB13H004"
    }
  ]
}