  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66070",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - USA including FL, VA, TX, NY, IL, AZ, MI, RI, NJ, SC, MO, TN, CA, MN, UT, GA, NE, OH, MD, PA, MT and internationally to Brazil, Australia, Canada, and Puerto Rico..",
      "recall_number": "Z-2274-2013",
      "product_description": "BD Epilor Syringe  BD Franklin Lakes, NJ 07417    BD EDC Laagstraat 57, B-9140 Temse-Belgium    The Epilor syringe is used in conjunction with an epidural needle with the purpose of finding the epidural space.",
      "product_quantity": "1,335,885 units",
      "reason_for_recall": "BD has received several reports indicating that the plunger of the 7 mL BD Epilor plastic BD Luer-Lok Loss of Resistance Syringe may stall or stick within the syringe preventing the user from detecting a loss of resistance upon entering the epidural space.",
      "recall_initiation_date": "20130823",
      "center_classification_date": "20130925",
      "termination_date": "20140319",
      "report_date": "20131002",
      "code_info": "510K K925902  D041521    Catalog Number: 405198  Lots  8310086  9246684  0035495  0053165  0056608  0056607  0096757  0120074  0126127  0152581  0242475  0258468  0258469  0258472  0291220  0291221  0291223  0291224  0278495  0291218  0291219  0291222  0291226  0291227  0291228  0336391  0336393  0336396  0336400  1005321  1006136  1033177  1033178  1094304  1209634  1210646  1213990  1213993  1276782  1307383  2032802  2033712  2060835  2065940  2084318  2121011  2122467  2205098  2212313  2276066  3038272  3042376  3063436  3063440  3085498  3085001  3085003  3085005  3123462  3123246  3184092              Catalog Number 405291  Lots  0154661  0174905  0252565  0279140  0301813  0347866  1012200  1035046  1063738  1095340  1124755  1180073  1213029  1262268  1273795  1312116  1320204  1356378  2041210  2058524  2083246  2135469  2158092  2170207  2194139  2205106  2223162  2223337  2250088  2286395  2303137  2303141  2334338  2334339  2355389  2355391  3057376  3120039        Catalog Number 405807  Lots  GD873422  GD874743  GD875419  GD876029  GD876912  GD879577  GD881128  GD883033        Catalog Number 405823  Lots  GD877068  GD882951  GD884312  GD885491  GD886192  GD887893  GD889527  B01F223D  B01F243D  B01G057D  B01G196D      Catalog Number 406078;   Lots  GD876003  GD877787  GD881169  GD883017  GD883660  GD885202  GD886283  GD888677  GD890442  B01F054D  B02F156D  B02F178D  B03F074D  B03F241D          Catalog number 406084  Lots  B03F090D  B06F326D  GD875369  GD880427  GD883025  GD888255  B05G087D          Catalog # 406119  Lots  GD873364  GD875625  GD877878  GD879015  GD881813  GD885129  GD889782  B02F051D  B02F096D  B01F135D  B03F222D    Catalog # 406121  Lots  GD887596  GD889618        Catalog # 406156  Lots  GD887778  GD889733  GD889980  GD890178  B02F214D  B02F292D  B02G157D  B03F292D  B03G158D  B04F238D  B05F241D"
    }
  ]
}