  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Jupiter",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62564",
      "recalling_firm": "Atlas Spine, Inc.",
      "address_1": "1555 Jupiter Park Dr Ste 4",
      "address_2": "N/A",
      "postal_code": "33458-8085",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "USA Distribution to the state of Florida only.",
      "recall_number": "Z-2274-2012",
      "product_description": "Apelo Modular Screwdriver labeled in part: \"APELO PEDICLE SCREW SYSTEM***Modular Screwdriver***Atlas Spine, Inc.***www.AtlasSpine.com*** 1555 Jupiter Park Drive #4***Jupiter, Florida 33458***561-741-1108***Tissue Sleeve***MFG Date: XX/XX/XXXX***Lot #: XXXX***Weldment***MFG Date: 03/23/2009***Sleeve Assembly***MFG Date: XX/XX/XXXX***Lot #: XXXX***Qty: 1 Rx Only***Non-Sterile. Sterilize Prior to Use*** Rev. A***.  The Modular Screwdriver  is used to insert modular pedicle screw implants which are part of the Apelo System.",
      "product_quantity": "4 pieces",
      "reason_for_recall": "Atlas Spine, Inc. recalled their Apelo Modular Screwdriver, a component of the Apelo Pedicle Screw System Instrumentation, because of complaints received about the tip breaking. If the tip is fractured inside of the modular screw, the tip could become implanted in the body or if the entire screw is removed, it could cause a delay in the surgery.",
      "recall_initiation_date": "20090604",
      "center_classification_date": "20120827",
      "termination_date": "20140716",
      "report_date": "20120905",
      "code_info": "PN# 70175-001  CAT#2006-41-0000  Lot # 00686"
    }
  ]
}