  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62578",
      "recalling_firm": "Biomet 3i, LLC",
      "address_1": "4555 Riverside Dr",
      "address_2": "N/A",
      "postal_code": "33410-4200",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-Distributed in the USA  and the countries to Spain, France, Germany, Switzerland, and Australia.",
      "recall_number": "Z-2271-2012",
      "product_description": "Low Profile Abutment, Non-Hexed Castable Cylinder Model LPCCC1.  Product is labeled in part:  \"***REF LPCCC1***Low profile Abutment Hexed Castable Cylinder***Cylinder; Cylindre; Zylinder; Cilindro; Cilindro; Cilindro***RX only***NON STERILE***Caution: Law prohibits dispensing without prescription***Consult instructions for use***    Castable cylinders are intended to be used to produce a definitive casting for dental restoration onto which porcelain can then be cast.",
      "product_quantity": "611 units",
      "reason_for_recall": "Biomet 3i conducted a recall on the device Low Profile Abutment, Non-Hexed Castable Cylinder due to some packages labeled LPCCC1, Hexed Castable Cylinder possibly containing LPCCC2, Non-Hexed Castable Cylinders.",
      "recall_initiation_date": "20120531",
      "center_classification_date": "20120824",
      "termination_date": "20131121",
      "report_date": "20120905",
      "code_info": "Lots: 2010100001, 2010080684, and 2010090191"
    }
  ]
}