      "recalling_firm": "Baxter Healthcare Corp.",
      "address_1": "Rt. 120 & Wilson Rd",
      "address_2": "N/A",
      "postal_code": "60073",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "45,865 units",
      "recall_number": "Z-2269-2012",
      "product_description": "Mini-Infuser Model 300XL Multispeed Infusion Pump, an Rx battery operated syringe infusion pump with an active time scale and rate selector switch;  product codes 2M8171 and 2M8171R.       Product Usage:  These product codes includes refurbished units. Intended for the controlled rate delivery of small volume parental fluids as prescribed by a physician.  Usage: The Mini-Infuser System is designed for controlled intermittent administration of intravenous medication.",
      "product_quantity": "19,380 units",
      "reason_for_recall": "Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum",
      "recall_initiation_date": "20110304",
      "center_classification_date": "20120824",
      "termination_date": "20120827",
      "report_date": "20120905",
      "code_info": "product codes 2M8171 and 2M8171R, all serial numbers"
    }
  ]
}