  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83377",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2268-2019",
      "product_description": "SOMATOM Definition AS, Model Number 8098027 - Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. These CT systems can be used for low dose lunch cancer screening in high risk populations (as defined by professional medical societies).",
      "product_quantity": "N/A",
      "reason_for_recall": "Potential for four safety-related software issues impacting systems utilizing syngo.CT software version VB10A which may result in a scan abort.",
      "recall_initiation_date": "20190703",
      "center_classification_date": "20190814",
      "termination_date": "20240510",
      "report_date": "20190821",
      "code_info": "Serial Numbers:  98474  92280  92261  65370  66032  66649  65299  92298  64516  92335  92300  66062  64218  92313  92296  92254  64051  92342  92277  92307  92336  92315  92288  98473  65145  65493  92323  92321  92243"
    }
  ]
}