{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Round Lake",
      "state": "IL",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62881",
      "recalling_firm": "Baxter Healthcare Corp.",
      "address_1": "Rt. 120 & Wilson Rd",
      "address_2": "N/A",
      "postal_code": "60073",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "45,865 units",
      "recall_number": "Z-2268-2012",
      "product_description": "Mini-Infuser Model 150XL Single Speed Infusion Pump, an Rx battery operated syringe infusion pump;       Product Usage:  These product codes includes refurbished units. Intended for the controlled rate delivery of small volume parental fluids as prescribed by a physician.  Usage: The Mini-Infuser System is designed for controlled intermittent administration of intravenous medication.",
      "product_quantity": "26,485 units",
      "reason_for_recall": "Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum",
      "recall_initiation_date": "20110304",
      "center_classification_date": "20120824",
      "termination_date": "20120827",
      "report_date": "20120905",
      "code_info": "product codes 2M8170 and 2M8170R, all serial numbers"
    }
  ]
}