  "results": [
    {
      "status": "Terminated",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65883",
      "recalling_firm": "Microgenics Corp",
      "address_1": "46360 Fremont Blvd",
      "address_2": "N/A",
      "postal_code": "94538-6406",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution including the states of  DC, MA, NJ, TX. FL, CA, WA, NY, TX, CT, KY, LA, AR, TN, AL, GA, CT, VA, NC, KS, MD, OH and IL",
      "recall_number": "Z-2267-2013",
      "product_description": "DRI Cocaine Metabolite Assay, labeled as Multigent Cocaine Assay for Abbott Laboratories.  100 ml, IVD    Mfg by:  Microgenics Corporation,  46360 Fremont Blvd.  Fremont, CA 94538    The DRI Cocaine Metabolite Enzyme Immunoassay is a homogeneous enzyme intended for the qualitative and semi-quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine with either 300 ng/mL or 150 ng/mL as a cutoff calibrator. The assay provided only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirm analytical result.",
      "product_quantity": "1473 units in distribution.",
      "reason_for_recall": "Surveillance of reagent stability for specific lots of  DRI Cocaine Metabolite Reagent show an unexpected decreasing change in the reaction rate, indicating a faster than expected reagent degradation.",
      "recall_initiation_date": "20130723",
      "center_classification_date": "20130923",
      "termination_date": "20131202",
      "report_date": "20131002",
      "code_info": "Part no. 0055, lot number 60053405, Exp 2/29/2016;  Part No. 0056, lot number 60053396, Exp 2/29/2016;  Part No. 10014593, lot number 60112588, Exp 2/29/2016; - not distributed  Part number #l40-20, lot number 24215M500, Exp 8/13/14."
    }
  ]
}