      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65992",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - All states in continental USA including DC except AK, AZ, RI  and WY.  and the countries of : AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, LEANON, MARINIQUE, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE,SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, LUXEMBOURG, PORTUGAL, VIETNAM and URUGUAY.",
      "recall_number": "Z-2266-2013",
      "product_description": "GE 1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant, HD23 Software and the following model numbers: 2395001, 2395001-2, 2395001-3, 2395001-4,  5269997-2, 5269997-4, 2377062-61, 5127452-3, 5127452, 5337894-3, 5337894-4, 5911000-4, 5911000-5, 53085735308573-2.    GE Discovery MR750w 3.0T, Discovery MR750 3.0T, Optima MR450w 1.5T, Discovery MR450 1.5T.    The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. IT is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including but not limited to head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used",
      "product_quantity": "1,246 (521 US and 725 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products.    For ClariView and Screen Save from GVTK (Gems Visualization ToolKit) Viewer, the MR Product can generate image headers with non-unique series identifiers, which when exported to some PACS systems, results in the series being associated wit",
      "recall_initiation_date": "20130409",
      "center_classification_date": "20130923",
      "termination_date": "20140505",
      "report_date": "20131002",
      "code_info": "REF 60850  Mfg Lot or Serial #  00000204882MR5  00000233303MR7  00000241103MR1  00000245171MR4  00000247204MR1  00000173038MR1  00000136104MR7  00000264553MR9  00000265096MR8  00000005532YR7  00000007097YR9  00000007556YR4  00000281841MR7  00000289583MR7  0000SGH919066P  00000013883YR4  0000SGH231PPQK  00002UA2050JKX  00002UA2400LWP  00002UA24315PT  00002UA24315PY  00002UA24315P5  00002UA24907W7  00000266797MRO  00000267189MR0  00000270550MR7  00000283466MR1  00000305180MR2  00000306054MR8  0000267533MR8A  Not Available  00000237000MR5  00000226580MR9  00000230105MR9  00000285956MR9  00000305386MR5  00000012562YR5  Not Available  000002UA4907WH  00002UA2400LWZ  00000177845MR5  00000165805MR3  00000836BD4DE2  00000012383YR6  00000269416MR4  Not Available  00002UA2400LWN  00000284254MR0  00000287469MR1  00000304797MR4  00002UA24907WK  00000005220YR9  00002UA2420VVZ  00002UA24907YC  00002UA24907YD  00002UA2510YTL  00000287996MR3  00000934BD5E4F  00000005966YR7  00000919BD5B8F  00000285151MR7  00000275969MR4  00000286019MR5  00000288669MR5  00001008BD6522  00000PH150024A  00000280423MR5  00000283613MR8  00000288416MR1  00000289840MR1  00000291157MR6  00000297685MR0  00000012064YR2  00000012526YR0  00000013660YR6  00000013820YR6  00000015654YR7  00001052XF502C  00001052XF5068  00001105XF5017  0000SGH048S36G  0000SGH105P174  0000SGH150RLYV  0000SGH150RLYX  0000SGH201TZ0V  0000SGH203QRPF  00002UA2050JKD  00002UA2050JL0  00002UA2070RGR  00002UA211170R  00002UA2281GGK  00002UA2281GGL  00002UA22827QF  00002UA2321LKW  00002UA2400LWJ  00002UA2400LWW  00002UA24907WJ  00000026639MR2  00000276261MR5  00000246260MR4  00000273065MR3  00000011594YR9  00000013110YR2  00001048XF5001  0000SGH106PMD7  814  00002UA2050JK7  00002UA2050JL2  00002UA2201G10  00002UA2201G19  00002UA22912Y5  00000264642MR0  00000305822MR9  00000265606MR4  Not Available  00000304307MR2  00000013981YR6  00002UA2400LX2  00000213294MR2  809  811  00000273256MR8  00000221025MR0  00000174670MR0  00000259576MR7  00000288072MR2  00000298886MR3  00000304463MR3  00000002761M65  00000002763M61  00000002766M64  00000003066M68  00000012720M69  00000012955M61  00001001BD631C  00001002BD6379  00001011BD6641  00001012BD66F7  00000PH150014A  00000PH150319A  00000PH150329A  00000PH150377A  00000283431MR5  00000284207MR8  00002UA9300HNM  00000012016YR2  00000013024YR5  00000013230YR8  00000013605YR1  00000013791YR9  00000013884YR2  00000014042YR6  00000014077YR2  00000015592YR9  00000016030YR9  00000292983MR4  00000296037MR5  00000296108MR4  00000298410MR2  00000298933MR3  00000299274MR1  00000299666MR8  00001018XF5033  00001018XF5035  00001027XF5008  00001037XF507E  00001037XF5078  00001042XF502A  00001042XF503B  00001043XF500A  00001043XF500C  00001043XF500D  00001043XF5011  00001046XF500C  00001046XF5016  00001046XF5018  00001046XF502D  00001047XF5001  00001048XF5011  00001048XF5016  00001049XF500E  00001049XF5028  00001052XF5008  00001101XF50A6  00001101XF50DA  00001101XF500D  00001102XF506C  00001102XF5077  00001105XF5003  00001105XF5010  00001105XF5026  0000SGH136SZ08  0000SGH136SZ1M  0000SGH139QLFS  0000SGH139QLGX  0000SGH139QLG0  0000SGH139QLHM  0000SGH201TZ0J  0000SGH203QRPG  0000SGH203QRQ7  0000SGH205RQ0Y  0000SGH212RCRM  0000SGH212RCR3  0000SGH231PPQH  0000SGH231PPQS  0000SGH242RN1D  0000SGH245TP29  0000SGH247Q8C3  00002UA2050JKB  00002UA2050JKN  00002UA2050JKR  00002UA2050JKT  00002UA2050JK3  00002UA2050JK4  00002UA2050JL5  00002UA2070RGX  00002UA211170M  00002UA2281GGJ  00002UA24907WB  00002UA2510YTF  00000299545MR4  00000265628MR8  00000266452MR2  00000267728MR4  00000267860MR5  00000268046MR0  00000270098MR7  00000270528MR3  00000294647MR3  00000297160MR4  00000297710MR6  00000297870MR8  00000298538MR0  00000298539MR8  00000298540MR6  00000298541MR4  00000298570MR3  00000299247MR7  00000299553MR8  00000300609MR5  00000300610MR3  00000300611MR1  00000300613MR7  00000300697MR0  00000302984MR0  00000302988MR1  00000303359MR4  00000303407MR1  00000303410MR5  00000303864MR3  00000303944MR3  00000304401MR3  00000304421MR1  00000304490MR6  00000304559MR8  00000304569MR7  00000304647MR1  00000304857MR6  00000305429MR3  00000305447MR5  00000305823MR7  00000265055MR4  00000265056MR2  00000298654MR5  00000303086MR3  00000305181MR0  00001051XF500B  0000SGH150RLY4  Not Available  00001140FMM0FU  Not Available  Not Available  Not Available  Not Available  Not Available  00002UA211170H  00000304248MR8  00000304750MR3  Not Available  00000PH150435A  00000016399M68  00000016658M67  Not Available  00002UA2321LKJ  00002UA4200LX1  00000241131MR2  00000261940MR1  00000007028M64  00002UA2050JKQ  00002UA22827QB  00002UA24907WF  00000269348MR9  00000200399MR4  00000012253YR1  00002UA22827QX  00002UA2400LWL  00000301918MR9  00000015792YR5  00001051XF5032  Not Available  Not Available  00000275465MR3  00000287379MR2  00000005432M60  00001043XF5019  00000PH150400A  Not Available  Not Available  Not Available  Not Available  00000013042YR7  00000272099MR3  00000283433MR1  00001052XF5021  0000SGH212RCRC  00002UA22827QJ  00000013142YR5  00002UA22912Y4  00000286730MR7  00000290035MR5  00002UA22827QS  00002UA22827QV  00000265845MR8  00000266189MR0  00000303354MR5  00000277200MR2  00002UA2050JKL  00000298690MR9  0000SGH136SZ17  00002UA21117OB  00000304774MR3  00000PH150389A    FMI 60842    Mfg Lot or Serial #  00000000UA0012  00000000UA0026  00000000UA0027  00000000UA0028  00000000UA0029  00000000UA0030  00000000UA0031  00000000UA0034  00000000UA0038  00000000UA0041  00000000UA0046  00000000UA0048  00000000UA0050  00000000UA0052  00000000UA0053  00000000UA0055  00000000UA0056  00000000UA0057  00000000UA0060  00000000UA0064  00000000UA0065  00000000UA0066  00000000UA0068  00000000UA0069  00000000UA0071  00000000UA0073  00000000UA0074  00000000UA0075  00000000UA0083  00000000UA0085  00000000UA0086  00000000UA0089  00000000UA0091  00000000UA0092  00000000UA0096  00000000UA0099  00000000UA0115  00000000UA0116  00000000UA0117  00000000UA0124  00000000UA0128  00000000UA0138  0000000UA0001A  0000000UA0011A  0000000UA0032A  00000029854MR5  00000030154MR0  00000030251MR2  00000123FMM0K4  00000288408MR8  00000288907MR9  00000289309MR7  00000289310MR5  00000289961MR5  00000290050MR4  00000290137MR9  00000290202MR1  00000290457MR1  00000290502MR4  00000290636MR0  00000290659MR2  00000290704MR6  00000290773MR1  00000290957MR0  00000291185MR7  00000291244MR2  00000291325MR9  00000291363MR0  00000291397MR8  00000291464MR6  00000291489MR3  00000291877MR9  00000291980MR1  00000292367MR0  00000292431MR4  00000292471MR0  00000292528MR7  00000292634MR3  00000292843MR0  00000293015MR4  00000293065MR9  00000293261MR4  00000293338MR0  00000293561MR7  00000293646MR6  00000293874MR4  00000293918MR9  00000294057MR5  00000294098MR9  00000294280MR3  00000294398MR3  00000294472MR6  00000294494MR0  00000294556MR6  00000294576MR4  00000294643MR2  00000294665MR5  00000294698MR6  00000294718MR2  00000294913MR9  00000294953MR5  00000294999MR8  00000295024MR4  00000295092MR1  00000295111MR9  00000295112MR7  00000295134MR1  00000295215MR8  00000295228MR1  00000295802MR3  00000296014MR4  00000296040MR9  00000296046MR6  00000296070MR6  00000296096MR1  00000296097MR9  00000296114MR2  00000296171MR2  00000296226MR0  00000296228MR0  00000296307MR2  00000296369MR2  00000296418MR7  00000296429MR4  00000296461MR7  00000296462MR5  00000296469MR0  00000296512MR7  00000296517MR6  00000296573MR9  00000296595MR2  00000296602MR6  00000296626MR5  00000296640MR6  00000296821MR2  00000296829MR5  00000296840MR2  00000296841MRO  00000296860MR0  00000296888MR1  00000296908MR7  00000297006MR9  00000297134MR9  00000297147MR1  00000297355MR0  00000297374MR1  00000297387MR3  00000297469MR9  00000297493MR9  00000297624MR9  00000297833MR6  00000297877MR3  00000297880MR7  00000297891MR4  00000297974MR8  00000298050MR6  00000298083MR7  00000298114MR0  00000298181MR9  00000298247MR8  00000298248MR6  00000298249MR4  00000298283MR3  00000298310MR4  00000298334MR4  00000298386MR4  00000298387MR2  00000298399MR7  00000298420MR1  00000298494MR6  00000298512MR5  00000298535MR6  00000298537MR2  00000298552MR1  00000298573MR7  00000298574MR5  00000298580MR2  00000298591MR9  00000298655MR2  00000298656MR0  00000298675MR0  00000298677MR6  00000298722MR0  00000298727MR9  00000298738MR6  00000298834MR3  00000298836MR8  00000298861MR6  00000298869MR9  00000298870MR7  00000298889MR7  00000298901MR0  00000298907MR7  00000298908MR5  00000298932MR5  00000298943MR2  00000298959MR8  00000298960MR6  00000298981MR2  00000299141MR2  00000299160MR2  00000299161MR0  00000299171MR9  00000299172MR7  00000299173MR5  00000299181MR8  00000299199MR0  00000299200MR6  00000299213MR9  00000299230MR3  00000299264MR2  00000299266MR7  00000299294MR9  00000299320MR2  00000299321MR0  00000299331MR9  00000299339MR2  00000299406MR9  00000299411MR9  00000299491MR1  00000299503MR3  00000299515MR7  00000299520MR7  00000299522MR3  00000299544MR7  00000299559MR5  00000299560MR3  00000299627MR0  00000299634MR6  00000299644MR5  00000299645MR2  00000299647MR8  00000299683MR3  00000299823MR5  00000299851MR6  00000299908MR4  00000299952MR2  00000299966MR2  00000299967MR0  00000300038MR7  00000300039MR5  00000300122MR9  00000300296MR1  00000300473MR6  00000300493MR4  00000300494MR2  00000300562MR6  00000300564MR2  00000300565MR9  00000300571MR7  00000300572MR5  00000300588MR1  00000300590MR7  00000300687MR1  00000300704MR4  00000300713MR5  00000300722MR6  00000300723MR4  00000300724MR2  00000300732MR5  00000300764MR8  00000300765MR5  00000300792MR9  00000300793MR7  00000300795MR2  00000300847MR1  00000300866MR1  00000300883MR6  00000300900MR8  00000300906MR5  00000300929MR7  00000300930MR5  00000300967MR7  00000300968MR5  00000300985MR9  00000300987MR5  00000301013MR9  00000301027MR9  00000301028MR7  00000301044MR4  00000301045MR1  00000301046MR9  00000301094MR9  00000301126MR9  00000301156MR6  00000301157MR4  00000301190MR5  00000301228MR3  00000301235MR8  00000301255MR6  00000301260MR6  00000301262MR2  00000301292MR9  00000301295MR2  00000301308MR3  00000301309MR1  00000301378MR6  00000301379MR4  00000301380MR2  00000301401MR6  00000301424MR8  00000301469MR3  00000301470MR1  00000301471MR9  00000301518MR7  00000301519MR5  00000301522MR9  00000301597MR1  00000301599MR7  00000301752MR2  00000301820MR7  00000301821MR5  00000301822MR3  00000301841MR3  00000301842MR1  00000301866MR0  00000301867MR8  00000301869MR4  00000301924MR7  00000301926MR2  00000301962MR7  00000302192MR0  00000302193MR8  00000302194MR6  00000302248MR0  00000302363MR7  00000302434MR6  00000302492MR4  00000302494MR0  00000302517MR8  00000302545MR9  00000302570MR7  00000302571MR5  00000302631MR7  00000302633MR3  00000302634MR1  00000302674MR7  00000302688MR7  00000302690MR3  00000302712MR5  00000302748MR9  00000302796MR8  00000302914MR7  00000302916MR2  00000302944MR4  00000302945MR1  00000302997MR2  00000302998MR0  00000303041MR8  00000303050MR9  00000303310MR7  00000303322MR2  00000303342MR0  00000303343MR8  00000303356MR0  00000303375MR0  00000303415MR4  00000303430MR3  00000303435MR2  00000303439MR4  00000303450MR1  00000303614MR2  00000303616MR7  00000303637MR3  00000303639MR9  00000303645MR6  00000303646MR4  00000303664MR7  00000303665MR4  00000303690MR2  00000303691MR0  00000303692MR8  00000303693MR6  00000303765MR2  00000303767MR8  00000303773MR6  00000303777MR7  00000303778MR5  00000303805MR6  00000303812MR2  00000303813MR0  00000303823MR9  00000303855MR1  00000303856MR9  00000303940MR1  00000303994MR8  00000304016MR9  00000304017MR7  00000304079MR7  00000304124MR1  00000304138MR1  00000304151MR4  00000304167MR0  00000304168MR8  00000304177MR9  00000304293MR4  00000304319MR7  00000304355MR1  00000304367MR6  00000304374MR2  00000304375MR9  00000304406MR2  00000304439MR3  00000304440MR1  00000304493MR0  00000304494MR8  00000304508MR5  00000304536MR6  00000304578MR8  00000304717MR2  00000304719MR8  00000304720MR6  00000304721MR4  00000304763MR6  00000304764MR4  00000304765MR1  00000304799MR0  00000304801MR4  00000304827MR9  00000304829MR5  00000304836MR0  00000304850MR1  00000304851MR9  00000304852MR7  00000304943MR4  00000304944MR2  00000305047MR3  00000305101MR8  00000305102MR6  00000305175MR2  00000305176MR0  00000305203MR2  00000305414MR5  00000305426MR9  00000305431MR9  00000305451MR7  00000305452MR5  00000305461MR6  00000305464MR0  00000305484MR8  00000305519MR1  00000305520MR9  00000305522MR5  00000305524MR1  00000305617MR3  00000305667MR8  00000305820MR3  00000305910MR2  000005148810-3  00000849BD53AB  00000903BD55CC  00000919BD5B82  00000944BD6172  00000950BD627C  00001001BD6325  00001009BD6598  00001012BD66BC  00001018XF503B  00001026XF5173  00001026XF5178  00001026XF5198  00001027XF5001  00001027XF500C  00001049XF5008  00001049XF5011  00001050XF5015  00001051XF5039  00001052XF5031  00001052XF5053  00001101XF50B7  00001101XF50B8  00001101XF50BC  00001102XF502F  00001102XF506B  00001105XF502C  00001106XF5016  00001106XF5046  00001106XF509C  00001106XF50A6  00001106XF50C0  00001106XF50CB  00001118FMM0WD  00001201FMM08B  00001203FMMOUJ  00001204FMM05D  00001219FMM0GT  00001221FMM184  00001236FMM01W  0000212825A73E  S0001106XF5025  00000299140MR4  00000306069MR6  00000300689MR7  00000300907MR3  00000304853MR5  00000903BD55D3  00000944BD6159  00001009BD65E7  00001011BD662B  00001101XF5037  00001106XF5035  00001106XF5052  00001106XF5078  00000000UA0102  00000299646MR0  00000300796MR0  00000000UA0023  00000301494MR1  00001134FMM107  00000000UA0039  00000291675MR7  00000293000MR6  00000293521MR1  00000293920MR5  00000295538MR3  00000295710MR8  00000296547MR3  00000298674MR3  00000300298MR7  00000301229MR1  00000303357MR8  00000304252MR0  00000304828MR7  00000000UA0022  00000293984MR1  00000296613MR3  00000298266MR8  00000299965MR4  00000303644MR9  00000303776MR9  00000296594MR5  00000300705MR1  00000300865MR3  00000301026MR1  00000301051MR9  00000301914MR8  00000303355MR2  00000304404MR7  00000305404MR6  00001101XF5070  00000000UA0079  00000298309MR6  00000296493MR0  00000296622MR4  00000303814MR8  00001105XF5047  00000288409MR6  00001140FMM0G4  00000295830MR4  00000297987MR0  00000304318MR9  00000000UA0080  00000000UA0095  00000292216MR9  00000293603MR7  00000301256MR4  00000303279MR4  00000304018MR5  00000304579MR6  00000304910MR3  00000294156MR5  00000301495MR8  00000299521MR5  00000000UA0043  00000301503MR9  00000302750MR5  00000298211MR4  00000299854MR0  00000301844MR7  00000000UA0112  00000297007MR7  00000298272MR6  00000298331MR0  00000299419MR2  00000300733MR3  00000300931MR3  00000301050MR1  00001009BD6582  00001018XF5029  00001026XF515A  00001101XF5039  00001101XF506C  00001102XF509B  00001204FMM04V  00000000UA0009  00000000UA0018  00000000UA0020  00000000UA0036  00000000UA0040  00000000UA0094  00000000UA0106  00000000UA0108  00000000UA0109  00000294943MR6  00000295180MR4  00000295486MR5  00000296560MR6  00000298566MR1  00000299147MR9  00000299293MR1  00000299330MR1  00000299407MR7  00000299455MR6  00000299545MR4  00000299546MR2  00000299853MR2  00000300734MR1  00000300791MR1  00000301194MR7  00000301458MR6  00000301577MR3  00000301586MR4  00000301726MR6  00000301925MR4  00000301970MR0  00000302493MR2  00000302516MR0  00000302689MR5  00000302915MR4  00000302920MR4  00000303054MR1  00000303852MR8  00000303854MR4  00001221FMM06N  2  00000000UA0063  00000291184MR0  00000291345MR7  00000293613MR6  00000294545MR9  00000295757MR9  00000296820MR4  00000297807MR0  00000299418MR4  00000299454MR9  00000300591MR5  00000301191MR3  00000302201MR9  00000303324MR8  00000304169MR6  00000304354MR4  00000304405MR4  00000305465MR7  00000000UA0024  00000000UA0025  00000000UA0087  00000300561MR8  00000301544MR3  00000291778MR9  00000294771MR1  00000296834MR5  00000301585MR6  00000288909MR5  00001203FMM0V5  00000000UA0051  00000300966MR9  00000300763MR0  00000293710MR0  00000297155MR4  00000000UA0021  00000000UA0037  00000000UAOO33  00000029804MR5  00000290456MR3  00000290545MR3  00000291274MR9  00000292973MR5  00000293206MR9  00000293642MR5  00000293724MR1  00000294332MR2  00000294967MR5  00000295709MR0  00000296321MR3  00000296514MR3  00000297132MR3  00000298102MR5  00000298493MR8  00000298501MR8  00000299340MR0  00000299684MR1  00000299685MR8  00000299825MR0  00000300712MR7  00000301199MR6  00000301296MR0  00000301872MR8  00000302675MR4  00000303280MR2  00000303309MR9  00000303311MR5  00000303638MR1  00000303694MR4  00000303766MR0  00000303822MR1  00000304078MR9  18  00000000UA0016  00000000UA0019  00000000UA0070  00000000UA0072  00000000UA0077  00000000UA0105  00000120FMM0VC  00000290985MR1  00000292844MR8  00000292928MR9  00000293042MR8  00000293327MR3  00000293919MR7  00000294575MR6  00000295847MR8  00000296032MR6  00000296141MR5  00000296227MR2  00000296332MR0  00000296499MR7  00000296814MR7  00000296879MR0  00000297164MR6  00000297468MR1  00000297494MR7  00000297636MR3  00000297701MR5  00000297715MR5  00000297861MR7  00000298073MR8  00000298242MR9  00000298385MR6  00000298430MR0  00000298567MR9  00000298734MR5  00000298850MR9  00000299341MR8  00000299464MR8  00000299633MR8  00000300589MR9  00000301125MR1  00000301236MR6  00000301291MR1  00000301520MR3  00000301523MR7  00000301775MR3  00000301868MR6  00000302544MR2  00000302711MR7  00000302996MR4  00000303323MR0  00000303376MR8  00000303414MR7  00000303449MR3  00000303774MR4  00000304077MR1  00000304123MR3  00000304295MR9  00000304538MR2  00000304555MR6  00000305616MR5  00001002BD6397  00000294397MR5  00000294713MR3  00000295485MR7  00000298431MR8  00000299453MR1  00000301912MR2  00000304492MR2  00000304580MR4  00000000UA0045  00000000UA0082  00000297154MR7  00000299452MR3  00000300452MR0  00000301432MR1  00000301496MR6  00000302432MR0  00000302795MR0  00000295827MR0  00000296291MR8  00000294548MR3  00000304800MR6  00000000UA0015  00000000UA0058  00000000UA0078  000001106XF039  00000288156MR3  00000292292MR0  00000293440MR4  00000295131MR7  00000298332MR8  00000298665MR1  00000298676MR8  00000299261MR8  00000301543MR5  00000301824MR9  00000301850MR4  00000302543MR4  00000302991MR5  00001102XF5054  00001117FMM0CP  00000297813MR8  00000298847MR5  00000305405MR3  00000297408MR7  00000000UA0035  00000000UA0049  00000294516MR0  00000297175MR2  00000297386MR5  00000297422MR8  00000298944MR0  00000299292MR3  00000301293MR7  00000301963MR5  00000301971MR8  00000303992MR2  00000305100MR0  00000000UA0010  00000000UA0042  00000030090MR1  00000291008MR1  00000291154MR3  00000294119MR3  00000294601MR0  00000295023MR6  00000296265MR2  00000297850MR0  00000298717MR0  00000298951MR5  00000299547MR0  00000299628MR8  00000299686MR6  00000300037MR9  00000300297MR9  00000300563MR4  00000300711MR9  00000300986MR7  00000303049MR1  00000303327MR1  00000303821MR3  00000304351MR0  00000304471MR6  00000304855MR0  00000305413MR7  00000305624MR9  00001002BD6387  00000298116MR5  00000300766MR3  00000300963MR6  00000301927MR0  00000302245MR6  00000303964MR1  00001003BD643B  00001009BD65A1  00001026XF518E  00001044XF5030  00001048XF500F  00001101XF5016  00001101XF5019  00001101XF5022  00001101XF5027  00001101XF5033  00001101XF5035  00001105XF502F  00001106XF502D  00001106XF5043  00001106XF504D  00001106XF5055  00001106XF507A  00001106XF508E  00001106XF50B4  00001106XF50C3  00001106XF50C5  00001106XF50D0  00001106XF50D7  00001106XF50DF  00001106XF5125  00001106XF512A  00000000UA0047  00000301158MR2  00000295060MR8  00000297156MR2  00000297898MR9  00000298115MR7  00000298251MR0  00000299824MR3  00000300472MR8  00000300474MR4  00000300586MR5  00000300848MR9  00000301197MR0  00000301502MR1  00000301584MR9  00000302203MR5  00000303048MR3  00000304438MR5  00000300587MR3  00000000UA0093  00000294451MR0  00000291429MR9  00000905BD566D  00001007BD646D  00001011BD633E  00001101XF50BB  00001102XF509F  00001106XF50E8  00001106XF50FF  00001117FMM0C7  00001134FMM10P  00001203FMM0VW  00001203FMM0W0  00001204FMM044  00003052527MR7  00000299229MR5  00000301047MR7"
    }
  ]
}