  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "West Chester",
      "address_1": "1302 Wrights Ln E",
      "reason_for_recall": "A screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the components within the set may be cytotoxic.",
      "address_2": "",
      "product_quantity": "4",
      "code_info": "Part Number: 145.17E Lot Number: EV01609, individual part numbers contained within 145.17E:   304.49  309.501  310.221  310.804  311.43  312.151  312.153  313.969  314.463  314.464  314.465  319.291  319.292  319.702  319.97  398.408  398.409  402.608-402.640  402.714-402.740  419.89  419.972  292.622, and   292.623.",
      "center_classification_date": "20140818",
      "distribution_pattern": "Distributed in the states of WI and MI.",
      "state": "PA",
      "product_description": "Synthes 3.0 MM TI Cannulated Screw Instrument and Implant Set",
      "report_date": "20140827",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Synthes, Inc.",
      "recall_number": "Z-2265-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68738",
      "termination_date": "20151218",
      "more_code_info": "",
      "recall_initiation_date": "20140630",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}