  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Duderstadt",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71715",
      "recalling_firm": "Otto Bock Healthcare GmbH",
      "address_1": "Max-Nader-Str. #15",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: AU, BE, CN, CZ, IT, NL, and SK.",
      "recall_number": "Z-2261-2015",
      "product_description": "Ottobock Pheon 3R62 Knee Joint    The 3R62 Pheon is a polycentric prosthetic knee joint intended to be used solely for lower limb prosthetic fittings. It is available in several versions to allow attachment to other components based on the patient need and physiology.",
      "product_quantity": "518 (439 US; 79 OUS)",
      "reason_for_recall": "Otto Bock Healthcare GmbH is voluntarily recalling all 3R62 Pheon polycentric prosthetic knee joints distributed since February 2014 because the locking function may fail under certain conditions which may result in a fall.",
      "recall_initiation_date": "20150716",
      "center_classification_date": "20150724",
      "termination_date": "20170621",
      "report_date": "20150805",
      "code_info": "The removal affects all devices manufactured and distributed since February 2014."
    }
  ]
}