  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Dubuque",
      "state": "IA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74043",
      "recalling_firm": "The Metrix Company",
      "address_1": "4400 Chavenelle Rd",
      "address_2": "N/A",
      "postal_code": "52002-2655",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom",
      "recall_number": "Z-2260-2016",
      "product_description": "Container fill and transfer sets, 250 mL, 500 mL, 1000 mL, 2000 mL, 3000 mL, or 4000 mL.",
      "product_quantity": "137,598 cases",
      "reason_for_recall": "An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.",
      "recall_initiation_date": "20160623",
      "center_classification_date": "20160722",
      "termination_date": "20230411",
      "report_date": "20160803",
      "code_info": "All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 58781, 58785, 62005, 62010, 62015, 62025, 62030, 62035, 64005, 64010, 64015, 64025, 64030, 64035, 64781, 65005, 65010, 65015, 65025, 65030, 65035, 66005, 66010, 66015, 66020, 66025, 66030, 66035, 66305, 66310, 66315, 66325, 66330, 66335, 66705, 66710, 66715, 66725, 66730, 66735, 66910, 66915, 66925, 66930, 66935, 67005, 67010, 67015, 67025, 67030, 67035, 69005, 69010, 69015, 69025, 69030, 69035"
    }
  ]
}