  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66126",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of : AL AZ, CA, CO, CT, FL, GA, IL, IN, MD,MI, MO, NJ, NY, NC, OK, PA, TX and WI.,  and the countries of :  AUSTRIA, BELGIUM, CANADA, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, IRAQ, JAPAN, LEBANON, NORWAY, PORTUGAL, SPAIN, SWEDEN and TAIWAN.",
      "recall_number": "Z-2257-2013",
      "product_description": "GE Healthcare, Seno Advantage, 5183685, 5183685-2, 5183685-3, 5183685-4, 5183685-5, 5183685-6, 5183685-7, 5183685-8, 5183685-9, 5342425, 5342425-2, 5342425-3, 5342425-4, 5342425-5.    Seno Advantage is a medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems. When interpreted by a trained physician, mammographic images displayed on the high-resolution monitors may be used as an element for diagnosis. Furthermore filmed images from all modalities may also be used as an element for diagnosis.",
      "product_quantity": "179",
      "reason_for_recall": "GE Healthcare has recently become aware that patient mismatch without a warning message could occur on Seno Advantage (SA) workstations under certain circumstances. This issue may impact patient safety.",
      "recall_initiation_date": "20120920",
      "center_classification_date": "20130919",
      "termination_date": "20160707",
      "report_date": "20130925",
      "code_info": "Seno Advantage Workstations with RIS CCOW integration: Patient  mismatch while receiving images - FMI 12187  Serial #  00000179275GE3  00000136608GE7  00000136609GE5  00000123803GE9  00000206492GE1  00000179274GE6  00000179273GE8  00000179272GE0  00000177331GE6  00000128528GE7  00000132314GE6  00000106422GE9  00000188371GE9  00000179281GE1  00000179282GE9  00000106426GE0  00000106427GE8  00000250274GE8  00000142837GE4  00000142836GE6  000000010001GS  00000114907GE9  00000158810GE2  00000084251GE8  00000112122GE7  00000123807GE0  000000310001GS  000000010008GS  000000010007GS  00000106420GE3  00000114906GE1  00000250277GE1  00000165231GE2  00000136610GE3  00000114088GE8  00000114089GE6  00000112926GE1  00000206490GE5  00000154938GE5  00000190211HP3  00000174027GE3  00000174026GE5  00000190278HP2  000000010004GS  00000164500GE1  00000158809GE4  00000158410GE1  00000165228GE8  00000165229GE6  00000123804GE7  00000166277GE4  00000114091GE2  00000106425GE2  00000106424GE5  00000179280GE3  00000200442GE2  00000250278GE9  00000126082GE7  00000206491GE3  00000164502GE7  00000164503GE5  000000510001GS  00000253622GE5  00000128529GE5  00000126080GE1  00000195402GE3  00000195403GE1  00000218675GE7  00000218674GE0  00000220618GE3  000000010011GS  00000190212HP1  00000220619GE1  000003510001GS  00000117141GE2  00000142832GE5  00000200440GE6  000003310002GS  00000022062GE9  000003310001GS  00000250290GE4  00000295750GE4  00000295751GE2  00000250276GE3  00000295753GE8  00000295752GE0  00000158812GE8  00000166279GE0  000000610001GS  00000075247GE7  00000106423GE7  00000080399GE9  000000610005GS  00000170113GE2  00000L02687311  000000ANT26269  000000G8087001  000000ANT23044  000000ANT25613  000000C7177002  00000193277GE1  00000149802GE1  00000149801GE3  00000165230GE4  000003510003GS  000000310002GS  000000310003GS  000004410002GS  000000610002GS  00000126078GE5  00000126079GE3  000000210002GS  00000164501GE9  00000117137GE0  00000119932GE2  00000142838GE2  000000Q7274004  000000M9335002  000000Q7274002  000000010005GS  000000N7302006  000000C7177009  000000C7177015  00000050606AW3  000000K7291002  000000Q7274005  000000C7177008  000000C7177010  00000Q11031001  00000Q11031003  000000610006GS  0000AG12200001  000000G9278001  00000L11224002  00000H11230001  0000ZA11210001  00000F11278001  0000AT11264001  000000S8340001  000000010001GS  00000J10019001  00000ZA9233001  00000N10193003  00000D10166001  000000B7339013  00000H11230002  0000AL11360001  00000H12052001  00000J11314001  000000S8135001  000000010003GS  000000T9021003  000000C7177020  000000010002GS  000000U7164006  00000050607AW1  000000Q7274001  00000054750AW5  00000054849AW5  000000010006GS  000000W9301001  000000T7250001  000000T7323003  000000T7323002  000000P8081001  000000S8135002  000000G8021002  00000050605AW5  00000050604AW8  00000G10264001  00000N10071001  00000P10046001  000000C7177019  0000AB11012002  00000M11006003  0000AB11012001  00000R11320001  000000R9058001  0000AT11264003",
      "more_code_info": ""
    }
  ]
}