      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85603",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide. Foreign distribution to Chile and Germany.",
      "recall_number": "Z-2256-2020",
      "product_description": "ProxiDiagnost N90, UDI 706100",
      "product_quantity": "91",
      "reason_for_recall": "The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo switch is not activated, the device may begin to generate smoke.",
      "recall_initiation_date": "20200423",
      "center_classification_date": "20200603",
      "termination_date": "20230330",
      "report_date": "20200610",
      "code_info": "Serial Numbers:    SN18000022  SN19000005  10000047  SN18000019  20000047  20000054  10000011  10000021  10000010  10000026  10000015  SN18000011  10000000  10000009  10000004  10000005  SN18000015  SN19000008  SN19000009  10000019  10000036  20000052  SN18000007  10000033  SN18000014  10000001  10000039  20000051  SN18000004  SN18000016  SN18000018  SN19000004  SN19000011  SN19000015  SN19000018  10000008  20000045  20000053  10000006  10000022  10000048    10000017  10000012  20000039  20000044  SN19000014  10000040  10000043  SN19000017  20000026  SN19000012  SN18000006  SN18000012  SN18000017  10000014  SN18000013  SN19000013  10000030  SN18000225  SN18000024  10000049  SN19000007  SN19000010  10000041  10000046  SN18000009  SN19000001  SN19000016  10000023  10000024  10000025  10000050  20000043  10000028  SN18000002  10000020  SN18000003  SN18000023  SN19000002  SN19000003  SN19000006  10000013  10000016  10000018  SN18000020   SN18000005  20000048"
    }
  ]
}