  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92615",
      "recalling_firm": "NeuMoDx Molecular Inc",
      "address_1": "1250 Eisenhower Pl",
      "address_2": "N/A",
      "postal_code": "48108-3281",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution to states of: Kansas, Florida, Texas, Ohio, New York, Michigan, Pennsylvania; and OUS (Foreign) to countries of: AT, BE, CH, DE, ES, FI, FR, GB, IT, LV, PL, RO, SE",
      "recall_number": "Z-2255-2023",
      "product_description": "NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems (NeuMoDx System(s)  Ref: 100100",
      "product_quantity": "20,016 cartridges",
      "reason_for_recall": "Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results",
      "recall_initiation_date": "20230524",
      "center_classification_date": "20230728",
      "report_date": "20230809",
      "code_info": "GTIN: 10814278020274  Lot Numbers:  117514, 117524, and 117528"
    }
  ]
}