  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Exeter",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88207",
      "recalling_firm": "Simpleware Product Group, SYNOPSYS NORTHERN EUROPE",
      "address_1": "Bradninch Hall",
      "address_2": "Castle Street",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: Nationwide Distribution in the states of  CA,  FL  and countries of : France, United Kingdom, Germany, Australia",
      "recall_number": "Z-2254-2021",
      "product_description": "Simpleware ScanIP medical software.  For transfer of imaging information from a medical scanner to an output file, and for simulating/evaluating surgical treatment options.",
      "product_quantity": "6 software's",
      "reason_for_recall": "A issue (bug) has been identified with the  interface and image software which could result in anatomical orientation tags/labels being displayed incorrectly. This could result in  operator misunderstanding and errors during resampling or re-alignment of image data",
      "recall_initiation_date": "20210604",
      "center_classification_date": "20210812",
      "termination_date": "20220929",
      "report_date": "20210818",
      "code_info": "Affected version/UDI:  Simpleware ScanIP Medical; Simpleware ScanIP O-2018.12: 00863520000338,  Simpleware ScanIP Medical; Simpleware ScanIP P-2019.09: 00863520000345,  Simpleware ScanIP Medical; Simpleware ScanIP Q-2020.03: 00863520000369,  Simpleware ScanIP Medical; Simpleware ScanIP Q-2020.06: 00863520000352,  Simpleware ScanIP Medical; Simpleware ScanIP R-2020.09: 00863520000376,  Simpleware ScanIP Medical; Simpleware ScanIP R-2021.03: 00863520000383"
    }
  ]
}