  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85689",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2254-2020",
      "product_description": "Multifire Endo Hernia 0 12mm Stapler with 4.8mm Disposable Loading Unit , Part No. 174025 - Product Usage: has application in endoscopic general surgery procedures for fixation of mesh in the repair of hernial defects and in other surgical specialties for approximation of tissue(s).",
      "product_quantity": "12,610 total",
      "reason_for_recall": "The product is recalled due to a manufacturing assembly error that may cause the staples to fail to properly advance within the staple cartridge when the instrument handle is compressed.",
      "recall_initiation_date": "20200512",
      "center_classification_date": "20200603",
      "report_date": "20200610",
      "code_info": "Lots  P9J1171Y                   P9K1395Y                   P9K1582Y"
    }
  ]
}