  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80209",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.",
      "recall_number": "Z-2253-2018",
      "product_description": "Apollo Anesthesia Machine; Cat. no. 8605310    Product Usage:  Indicated as a continuous flow anesthesia system. The Apollo may be used for manually assisted or automatic ventilation, delivery of gases and anesthetic vapor, and monitoring of oxygen and CO2 concentration, breathing pressure, respiratory volume, and anesthetic agent concentration and identification. It is an inhalation anesthesia machine for use in operating, induction, and recovery rooms. It can be used with rebreathing systems, semiclosed to virtually closed systems with low flow and minimal flow techniques, and non-rebreathing systems (with the Auxiliary Common Gas Outlet). It may be used with O2, N2O, and Air supplied by a medical gas pipeline system or by externally mounted gas cylinders. Anesthetic agent can be delivered via vaporizers mounted on the machine. The Apollo is equipped with a compact breathing system, providing fresh-gas decoupling, PEEP, and pressure limitation. It has an electrically driven and electronically controlled ventilator.",
      "product_quantity": "72",
      "reason_for_recall": "The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result  in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.",
      "recall_initiation_date": "20180524",
      "center_classification_date": "20180619",
      "termination_date": "20200512",
      "report_date": "20180627",
      "code_info": "ASKM-0152 ASKN-0138  ASKM-0193 ASKN-0139  ASKM-0202 ASKN-0143  ASKM-0203 ASKN-0144  ASKM-0211 ASKN-0148  ASKM-0212 ASKN-0150  ASKM-0213 ASKN-0151  ASKM-0214 ASKN-0152  ASKM-0221 ASKN-0163  ASKM-0222 ASKN-0164  ASKM-0227 ASKN-0165  ASKM-0228 ASLA-0001  ASKM-0246 ASLA-0002  ASKM-0269 ASLA-0003  ASKM-0272 ASLA-0092  ASKM-0328 ASLA-0121  ASKM-0332 ASLA-0124  ASKM-0333 ASLA-0131  ASKM-0337 ASLA-0132  ASKM-0339 ASLA-0145  ASKM-0340 ASLA-0146  ASKM-0360 ASLA-0152  ASKM-0362 ASLA-0153  ASKM-0366 ASLA-0154  ASKM-0370 ASLA-0163  ASKM-0371 ASLA-0164  ASKM-0408 ASLA-0165  ASKM-0409 ASLA-0166  ASKM-0415 ASLA-0173  ASKN-0012 ASLA-0174  ASKN-0035 ASLA-0175  ASKN-0039 ASLA-0179  ASKN-0040 ASLA-0180  ASKN-0113 ASLA-0181  ASKN-0114 ASLA-0182  ASKN-0116 ASLA-0183"
    }
  ]
}