  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irving",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85449",
      "recalling_firm": "Abbott Laboratories",
      "address_1": "1915 Hurd Dr",
      "address_2": "N/A",
      "postal_code": "75038",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of  NC, ND, SC, TN.    The countries of  United Kingdom, France, New Zealand, Pakistan, Turkey.",
      "recall_number": "Z-2251-2020",
      "product_description": "Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation.",
      "product_quantity": "11",
      "reason_for_recall": "In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.",
      "recall_initiation_date": "20200402",
      "center_classification_date": "20200603",
      "termination_date": "20210428",
      "report_date": "20200610",
      "code_info": "All serial numbers in the field"
    }
  ]
}