{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80209",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.",
      "recall_number": "Z-2251-2018",
      "product_description": "Fabius GS Tiro Anesthesia Machine; Cat. no. 8606000    Product Usage:  Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Dr¿ger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less",
      "product_quantity": "31",
      "reason_for_recall": "The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result  in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.",
      "recall_initiation_date": "20180524",
      "center_classification_date": "20180619",
      "termination_date": "20200512",
      "report_date": "20180627",
      "code_info": "ASKM-0110  ASKM-0111  ASKM-0196  ASKM-0209  ASKM-0232  ASKM-0244  ASKM-0246  ASKN-0032  ASKN-0033  ASKN-0034  ASKN-0037  ASKN-0041  ASKN-0043  ASKN-0068  ASKN-0093  ASLA-0009  ASLA-0021  ASLA-0023  ASLA-0031  ASLA-0032  ASLA-0123  ASLA-0130  ASLA-0135  ASLA-0136  ASLA-0147  ASLA-0155  ASLA-0158  ASLA-0163  ASLA-0166  ASLA-0170  ASLA-0195"
    }
  ]
}