  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77220",
      "recalling_firm": "Alphatec Spine, Inc.",
      "address_1": "5830 El Camino Real",
      "address_2": "N/A",
      "postal_code": "92008-8816",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of (TX, AL, UT, FL, IA, SD, NV, OH, CA, MN, CO, AZ, OK, MI, VA, KY, and Puerto Rico) and country of Japan",
      "recall_number": "Z-2249-2017",
      "product_description": "Arsenal Spinal Fixation System, Set Screw, Part No. 47027      Product Usage:  The Arsenal Spinal Fixation System is intended for posterior, non-cervical, spinal fixation as an adjunct to fusion for the treatment of degenerative disease, deformity, and trauma indications. The Arsenal System consists of a variety of shapes and sizes of rods, screws, hooks, connectors, and bridges that provide temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. The screws, hooks, connectors, and bridges are manufactured from surgical grade titanium alloy (Ti-6AI-4V ELl). The rods are available in commercially pure titanium, titanium alloy, and cobalt chrome (CP Ti Grade 4, Ti-6AI-4V ELl, and Co-28Cr-6Mo).",
      "product_quantity": "16,959 units",
      "reason_for_recall": "Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw  postoperative disengagement from the screw body.",
      "recall_initiation_date": "20170505",
      "center_classification_date": "20170606",
      "termination_date": "20170606",
      "report_date": "20170614",
      "code_info": "671949 681640  684646 687843 690206 692720 693064  671948 681641 684646  687843  690206 692705  674136 681642 684647 687847 690210 692718  674137 682061 686374 687847 691848S1 692719  674431 682060 684645 689847 691848  692855  674432  682062 686372 689847 691849  692855  674433 682063 686375 689849 692597 692706  671950 684429 687839 690127 692597 692864  674429 684430 687839 689848  692364 692910  674430 684432 687840 690207 692598 692911  681638S1 684644 687840 690207 691590 692911  681638 684433 687842 690208 692599 693065  681639  684431 687845 690209 692704 693057"
    }
  ]
}