{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "Zimmer is recalling various trauma screws that failed a leak test due to damage of the inner cavity only.",
      "address_2": "",
      "product_quantity": "47,682 units total",
      "code_info": "Item Number 47-4865-030, 035, 040, 045, 050, 055, 060, 065, 070, 075, 080, 085, 090, 100-00",
      "center_classification_date": "20140815",
      "distribution_pattern": "National Distribution: AK, AL, AZ, CA, CT, DC, FL, GA, US VIRGIN ISLANDS, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WI, WV, WY  Foreign Distribution: Canada, Mexico, ARGENTINA, AUSTRALIA, BELGIUM,BRAZIL, CHINA, EL SALVADOR, ENGLAND, GERMANY, INDIA, ITALY, JAPAN, JORDAN, KOREA, NEW ZEALAND, SAUDI ARABIA, SINGAPORE, SPAIN, SWITZERLAND, TAIWAN, UAE",
      "state": "IN",
      "product_description": "6.5 mm Cancellous Screw Fully Threaded   Sterile  zimmer",
      "report_date": "20140827",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-2249-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68819",
      "termination_date": "20150618",
      "more_code_info": "",
      "recall_initiation_date": "20140722",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}