  "results": [
    {
      "status": "Terminated",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88266",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "15 Hampshire St Bldg 5",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2244-2021",
      "product_description": "Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811).  Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate.    Product Code: 43201",
      "product_quantity": "7197 units",
      "reason_for_recall": "The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.",
      "recall_initiation_date": "20210616",
      "center_classification_date": "20210816",
      "termination_date": "20230104",
      "report_date": "20210825",
      "code_info": "All lot codes"
    }
  ]
}