  "results": [
    {
      "status": "Terminated",
      "city": "Fremont",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71608",
      "recalling_firm": "EXP Pharmaceutical Services Corp",
      "address_1": "48021Warm Springs Boulevard",
      "address_2": "N/A",
      "postal_code": "94539",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in DC and the states of CO, CA, FL, TX, and AR.",
      "recall_number": "Z-2244-2015",
      "product_description": "Medical Device Exchange  SECURESTRAP 5mm Absorbable Strap Fixation Device;  Pouch, Tyvek/Mylar, Sterile, Single Use Only;  OEM Catalog number STRAP25;  EXP (repackager), Ethicon (OEM)    Strap, implantable.",
      "product_quantity": "2",
      "reason_for_recall": "EXP did not register as a medical device establishment, list the devices being recalled, or establish its own quality system and instead relied on the fact that its vendors were registered and had their own quality systems.",
      "recall_initiation_date": "20150211",
      "center_classification_date": "20150724",
      "termination_date": "20151230",
      "report_date": "20150805",
      "code_info": "OEM Catalog number  STRAP25."
    }
  ]
}