{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warwick",
      "state": "RI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92213",
      "recalling_firm": "Davol, Inc.",
      "address_1": "100 Crossings Blvd",
      "address_2": "N/A",
      "postal_code": "02886-2850",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.",
      "recall_number": "Z-2243-2023",
      "product_description": "Ventralight ST w Echo PS 4.5\" Circle   A low profile, bioresorbable, coated, permanent mesh, with a pre-attached removable positioning system, designed for the reconstruction of soft tissue deficiencies during laparoscopic ventral hernia  repair.   Catalog Number: 5955450G (EU only)",
      "product_quantity": "169 units",
      "reason_for_recall": "Balloon may not inflate as intended. If the balloon  does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection",
      "recall_initiation_date": "20230425",
      "center_classification_date": "20230727",
      "report_date": "20230802",
      "code_info": "UDI-DI: (01)00801741201950(17)240528(10) Lot Number HUGT1609"
    }
  ]
}