{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 GBC Drive, Mailstop 514",
      "reason_for_recall": "When using the EasyLink Quality Control Package the system may not perform as intended for Multi-Rule QC violations [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions send with the Urgent Field Safety Notice.  When this situation occurs, QC status may not be flagged appropriately.  If auto-verification rules are defined to hold patient results when QC status is out, the patient",
      "address_2": "PO BOX 6101",
      "product_quantity": "637",
      "code_info": "Part Number:  1000034805, Version 5.0 and Version 5.0 Service Pack 1 through 4",
      "center_classification_date": "20120822",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Canada, France, Germany, Spain, Portugal, Italy, Japan, Norway, Australia, Austria, New Zealand, Belgium, Switzerland, Netherlands, Malaysia, Slovenia, South Korea, Denmark, Saudi Arabia, Slovakia, Taiwan, Philippines, Czech Republic, and Singapore.",
      "state": "DE",
      "product_description": "EasyLink Informatic System is included with the Dimension Vista System.  The Dimension Vista Integrated System is an in vitro diagnostic device intended tot duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.    The EasyLink Informatics System is classified as a data calculator and processing module for use with in vitro diagnostic devices. The EasyLink is included with the Dimension Vista System and may be used with other Siemens analyzers as a communications and connectivity workstation for integration with laboratory information system (LIS) networks.",
      "report_date": "20120829",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2242-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62460",
      "termination_date": "20140410",
      "recall_initiation_date": "20120529",
      "postal_code": "19714-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}