  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74576",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) Internationally  to Australia, The Netherlands, and Japan",
      "recall_number": "Z-2240-2016",
      "product_description": "TSP Hip Fracture Plating System  device, fixation, proximal femoral, implant  The Biomet Hip Fracture Plating Systems are indicated for use in the open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures and intertrochanteric fractures. Fracture plates with six or more femoral shaft fixation screw holes are also indicated for subtrochanteric fractures as well as proximal femoral osteotomies",
      "product_quantity": "5458 units",
      "reason_for_recall": "The screw may fracture at the junction of the shaft and the head during implantation. The patient may retain a foreign body if the fractured screw cannot be removed. Soft tissue irritation will not occur as the retained portion of the screw is below the surface of the plate. A delay in surgery less than 30 minutes may occur.",
      "recall_initiation_date": "20160610",
      "center_classification_date": "20160719",
      "termination_date": "20170720",
      "report_date": "20160727",
      "code_info": "All Lots.    Part Number / Description:    14-451002  130 DEG 2-HOLE LEFT NECK FX  14-451004  130 DEG 4-HOLE LEFT NECK FX  14-451070  7.5MM TELESCOPE LAG SCR 70MM  14-451075  7.5MM TELESCOPE LAG SCR 75MM  14-451080  7.5MM TELESCOPE LAG SCR 80MM  14-451085  7.5MM TELESCOPE LAG SCR 85MM  14-451090  7.5MM TELESCOPE LAG SCR 90MM  14-451095  7.5MM TELESCOPE LAG SCR 95MM  14-451100  7.5MM TELESCOPE LAG SCR 100MM  14-451102  130 DEG 2-HOLE RIGHT NECK FX  14-451104  130 DEG 4-HOLE RIGHT NECK FX  14-451105  7.5MM TELESCOPE LAG SCR 105MM  14-451110  7.5MM TELESCOPE LAG SCR 110MM  14-451115  7.5MM TELESCOPE LAG SCR 115MM  14-451120  7.5MM TELESCOPE LAG SCR 120MM  14-451125  7.5MM TELESCOPE LAG SCR 125MM  14-451130  7.5MM TELESCOPE LAG SCR 130MM  14-451500  ANGLE GUIDE 130 DEG  14-451502  GUIDE WIRE 3.0MM  14-451503  GUIDE PIN DRIVER 3.0MM  14-451504  TARGETING GUIDE 130 DEG LEFT  14-451505  TARGETING GUIDE 130 DEG RIGHT  14-451509  JIG ATTACHMENT DRIVER 3.5MM  14-451510  NON-LOCK GUIDE SLEEVE 7.5MM  14-451511  LOCKING GUIDE SLEEVE 7.5MM  14-451512  ENTRY TROCAR 7.5 MM  14-451513  DRILL BIT 3.0MM  14-451525  GUIDE WIRE DEPTH GAUGE 3.0MM  14-451526  STEP DRILL CANNULATED 7.5MM  14-451527  STEP DRILL NON-CANN 7.5MM  14-451528  TSP HIP FX TAP 7.5MM  14-451529  HEX DRIVER NECK SLEEVE 5.0MM  14-451530  DRIVER CONNECTOR SHAFT 5.0MM  14-451531  HEX DRIVER NECK SHAFT 3.5MM  14-451532  INSERTER LIMITED COLLAPSE  14-451533  DEPTH GAUGE LIMITED COLLAPSE  14-451535  SCREW GUIDE SLEEVE 4.5MM  14-451536  TROCAR 4.5MM SCREW  14-451537  DRILL GUIDE LOCKING 3.5MM  14-451538  DRILL GUIDE NON-LOCK 3.5MM  14-451539  DRILL BIT 3.5MM  14-451540  SCREW DEPTH GAUGE 4.5MM  14-451541  SCREW DEPTH GAUGE HOOK 4.5MM  14-451542  TSP HIP FX CORTICAL TAP 4.5MM  14-451543  DRIVER 5.0 HEX/3.5 HEXALOBE  14-451544  CONNECTOR SHAFT 4.5MM SCREW  14-451546  T HANDLE W/ HUDSON  14-451550  IMPACTION HANDLE  14-451551  IMPACTION GUIDE PAD  14-451552  IMPACTION PLATE PAD  14-451554  GUIDE WIRE ALIGNMENT ARM  14-451560  GUIDE WIRE PUSHER 2.8MM  14-451561  GUIDE WIRE PLACEMENT TROCAR  14-451562  GUIDE WIRE PLACEMENT SLEEVE  14-451565  SCREW EXTRACTOR 7.5MM SCREW  14-451567  SCREW EXTRACTOR 4.5MM SCREW  14-451568  LAG SCREW REMOVAL TOOL 7.5MM  14-451570  BONE HOOK  14-451571  BALL SPIKE PUSHER  14-451573  REDUCTION TOOL  14-451580  NECK FX INSTRUMENT CASE  14-453005  LIMITED COLLAPSE SLEEVE 5MM  14-453010  LIMITED COLLAPSE SLEEVE 10MM  14-453015  LIMITED COLLAPSE SLEEVE 15MM  14-453020  LIMITED COLLAPSE SLEEVE 20MM  14-454014  4.5MM CORT LOCK SCREW 14MM  14-454016  4.5MM CORT LOCK SCREW 16MM  14-454018  4.5MM CORT LOCK SCREW 18MM  14-454020  4.5MM CORT LOCK SCREW 20MM  14-454022  4.5MM CORT LOCK SCREW 22MM  14-454024  4.5MM CORT LOCK SCREW 24MM  14-454026  4.5MM CORT LOCK SCREW 26MM  14-454028  4.5MM CORT LOCK SCREW 28MM  14-454030  4.5MM CORT LOCK SCREW 30MM  14-454032  4.5MM CORT LOCK SCREW 32MM  14-454034  4.5MM CORT LOCK SCREW 34MM  14-454036  4.5MM CORT LOCK SCREW 36MM  14-454038  4.5MM CORT LOCK SCREW 38MM  14-454040  4.5MM CORT LOCK SCREW 40MM  14-454042  4.5MM CORT LOCK SCREW 42MM  14-454044  4.5MM CORT LOCK SCREW 44MM  14-454046  4.5MM CORT LOCK SCREW 46MM  14-454048  4.5MM CORT LOCK SCREW 48MM  14-454050  4.5MM CORT LOCK SCREW 50MM  14-454052  4.5MM CORT LOCK SCREW 52MM  14-454054  4.5MM CORT LOCK SCREW 54MM  14-454056  4.5MM CORT LOCK SCREW 56MM  14-454058  4.5MM CORT LOCK SCREW 58MM  14-454060  4.5MM CORT LOCK SCREW 60MM  14-454065  4.5MM CORT LOCK SCREW 65MM  14-454070  4.5MM CORT LOCK SCREW 70MM  14-454075  4.5MM CORT LOCK SCREW 75MM  14-454080  4.5MM CORT LOCK SCREW 80MM  14-454085  4.5MM CORT LOCK SCREW 85MM  14-454090  4.5MM CORT LOCK SCREW 90MM  14-454095  4.5MM CORT LOCK SCREW 95MM  14-455014  4.5MM CORT NON-LOCK SCREW 44MM  14-455016  4.5MM CORT NON-LOCK SCREW 16MM  14-455018  4.5MM CORT NON-LOCK SCREW 18MM  14-455020  4.5MM CORT NON-LOCK SCREW 20MM  14-455022  4.5MM CORT NON-LOCK SCREW 22MM  14-455024  4.5MM CORT NON-LOCK SCREW 24MM  14-455026  4.5MM CORT NON-LOCK SCREW 26MM  14-455028  4.5MM CORT NON-LOCK SCREW 28MM  14-455030  4.5MM CORT NON-LOCK SCREW 30MM  14-455032  4.5MM CORT NON-LOCK SCREW 32MM  14-455034  4.5MM CORT NON-LOCK SCREW 34MM  14-455036  4.5MM CORT NON-LOCK SCREW 36MM  14-455038  4.5MM CORT NON-LOCK SCREW 38MM  14-455040  4.5MM CORT NON-LOCK SCREW 40MM  14-455042  4.5MM CORT NON-LOCK SCREW 42MM  14-455044  4.5MM CORT NON-LOCK SCREW 44MM  14-455046  4.5MM CORT NON-LOCK SCREW 46MM  14-455048  4.5MM CORT NON-LOCK SCREW 48MM  14-455050  4.5MM CORT NON-LOCK SCREW 50MM  14-455052  4.5MM CORT NON-LOCK SCREW 52MM  14-455054  4.5MM CORT NON-LOCK SCREW 54MM  14-455056  4.5MM CORT NON-LOCK SCREW 56MM  14-455058  4.5MM CORT NON-LOCK SCREW 58MM  14-455060  4.5MM CORT NON-LOCK SCREW 60MM  14-455065  4.5MM CORT NON-LOCK SCREW 65MM  14-455070  4.5MM CORT NON-LOCK SCREW 70MM  14-455075  4.5MM CORT NON-LOCK SCREW 75MM  14-455080  4.5MM CORT NON-LOCK SCREW 80MM  14-455085  4.5MM CORT NON-LOCK SCREW 85MM  14-455090  4.5MM CORT NON-LOCK SCREW 90MM  14-455095  4.5MM CORT NON-LOCK SCREW 95MM",
      "more_code_info": ""
    }
  ]
}