  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74639",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) Netherlands, Canada, Japan, Switzerland, and United Kingdom",
      "recall_number": "Z-2238-2016",
      "product_description": "2.3MM Tapered Router,  Product Number 5407-FA2-023  A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and Throat  (ENT)/Otology /Neurotology/ Otorhinolaryngology; Craniofacial (bones of the skull and supraorbital region); and  Sternotomy. They are intended to be used with the Stryker CORE\" system.",
      "product_quantity": "3,163",
      "reason_for_recall": "A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be  more susceptible to breakages.  Risk to Health:  Injury to critical soft tissue requiring surgical intervention may occur.",
      "recall_initiation_date": "20160624",
      "center_classification_date": "20160719",
      "termination_date": "20170323",
      "report_date": "20160727",
      "code_info": "Lot numbers:  15077027, 15077067, 15077097, 15078087, 15105017, 15108017, 15108047, 15108077, 15108087, 15108097, 15146017, 15146027, 15146037, 15146047",
      "more_code_info": ""
    }
  ]
}