{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85643",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US nationwide including US Virgin Islands,   Australia,   Austria,   Bahrain,   Belgium,   Bermuda,   Bulgaria,   Burundi,   Canada,   Chile,   China,   Croatia,   Czech Republic,   Denmark,   Dominican Republic,   Ecuador,   Egypt,   El Salvador,   Estonia,   Finland,   France,   Germany,   Greece,   Guernsey,   Hong Kong,   India,   Ireland,   Israel,   Italy,   Japan,   Jersey,   Jordan,   Kazakhstan,   Republic of Korea,   Kuwait,   Latam,   Latvia,   Lithuania,   Macedonia,   Malaysia,   Mexico,   Morocco,   Netherlands,   New Zealand,   Norway,   OMAN,   State of Palestine,    Panama,   Peru,   Poland,   Portugal,   Qatar,   Romania,   Russia,   Saudi Arabia,   Singapore,   Slovakia,   Slovenia,   South Africa,   Spain,   Sweden,   Switzerland,   Tunisia,   Turkey,   Ukraine,   United Arab Emirates,   United Kingdom,   Uruguay,   Venezuela,   and Yemen",
      "recall_number": "Z-2236-2020",
      "product_description": "Giraffe Incubator with installed Servo Oxygen module. Infant incubator.",
      "product_quantity": "5,883 units total",
      "reason_for_recall": "Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.",
      "recall_initiation_date": "20200428",
      "center_classification_date": "20200615",
      "termination_date": "20240923",
      "report_date": "20200624",
      "code_info": "Giraffe Incubator: serial numbers HDGA through HDGQ;      Servo Oxygen Board spare part kit: 6600-0233-850;      Servo Oxygen upgrade kits: 6600-0678-800, 6600-0678-801, 6600-0678-802, 6600-0678-803, 66000849-800, 6600-0849-801, 6600-0849-802, 6600-0849-803, 6600-0866-800, 6600-0866-801, 6600-0867800, 6600-0867-801, M1142293 or M1142295.",
      "more_code_info": ""
    }
  ]
}