  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66037",
      "recalling_firm": "Smith & Nephew, Inc. Endoscopy Division",
      "address_1": "150 Minuteman Drive",
      "address_2": "N/A",
      "postal_code": "01810-1031",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide)  Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.",
      "recall_number": "Z-2235-2013",
      "product_description": "TWINFIX\" Ultra PLLA/HA 6.5 mm with 2 ULTRABRAID\" Suture, White / Black: Part Number: 72202608    Biodegradable suture anchor",
      "product_quantity": "1308 US",
      "reason_for_recall": "Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.",
      "recall_initiation_date": "20130806",
      "center_classification_date": "20130917",
      "termination_date": "20160420",
      "report_date": "20130925",
      "code_info": "50325164  50326837  50336721  50336771  50339645  50339841  50341067  50341975  50344394  50349837  50350415  50351794  50359331  50363876  50368160  50370201  50372046  50375166  50377788  50378916  50380676  50381053  50388605  50392271  50392403  50395799  50403938  50411076  50417379  50422358  50433010  50433121  50438323  50448200  50460082",
      "more_code_info": ""
    }
  ]
}