  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80193",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.",
      "recall_number": "Z-2233-2018",
      "product_description": "BALL TIP GUIDE WIRE 3.2MMX55CM; 469055     Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.",
      "product_quantity": "1234",
      "reason_for_recall": "The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.",
      "recall_initiation_date": "20180516",
      "center_classification_date": "20180615",
      "termination_date": "20240405",
      "report_date": "20180627",
      "code_info": "410640  410650  413650  561810  700980  828080  843730  244860  579030  265412  276373  562990  754210  235170  546830  579480  929310  013900  958060  560420  560450  C32280  C32290  E36750  056530  241100  089550  528280  015770  564990  643870  067590  822470  822480  670840  038710  391720  380280  558070  016760  174930  695650  399280  932230  963110"
    }
  ]
}