  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Marlborough",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92570",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "100 Boston Scientific Way",
      "address_2": "N/A",
      "postal_code": "01752-1234",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Canada.",
      "recall_number": "Z-2232-2023",
      "product_description": "Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy  (Box of 5)UPN: M006L8406911",
      "product_quantity": "205 units",
      "reason_for_recall": "Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched",
      "recall_initiation_date": "20230525",
      "center_classification_date": "20230721",
      "report_date": "20230802",
      "code_info": "GTIN: 08714729978725 Lot Numbers:  30521480, 30521485 ,30521486 Exp. Date:  20-OCT-2024 to 10-Nov-2025"
    }
  ]
}