      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85613",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution.",
      "recall_number": "Z-2225-2020",
      "product_description": "Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 7.5X45MM, Item Number 810M7545 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.",
      "product_quantity": "3235 total",
      "reason_for_recall": "There are reports of inadequate locking of the extended tab screws.  Complaints were reported where the closure top backed out of the screw post-operatively.",
      "recall_initiation_date": "20200505",
      "center_classification_date": "20200602",
      "termination_date": "20220715",
      "report_date": "20200610",
      "code_info": "UDI: (01)00887868325281; Lot Numbers: C19I0005  C19I0016  C19J0018  C19L0010  C20A0009  C20A0011  C20A0013  SBM108415  SBM118997  SBM120813  SBM122160"
    }
  ]
}