  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98945",
      "recalling_firm": "Fresenius Kabi USA, LLC",
      "address_1": "50 High St Ste 50",
      "address_2": "",
      "postal_code": "01845-2620",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;",
      "recall_number": "Z-2224-2026",
      "product_description": "Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004",
      "product_quantity": "18,444",
      "reason_for_recall": "Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.",
      "recall_initiation_date": "20260506",
      "center_classification_date": "20260529",
      "report_date": "20260610",
      "code_info": "Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:",
      "more_code_info": ""
    }
  ]
}