  "results": [
    {
      "status": "Terminated",
      "city": "Princeton",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88312",
      "recalling_firm": "Integra LifeSciences Corp.",
      "address_1": "1100 Campus Rd",
      "address_2": "N/A",
      "postal_code": "08540-6650",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution - Foreign Distribution: Hong Kong",
      "recall_number": "Z-2224-2021",
      "product_description": "CODMAN RANEY SCALP CLIPS-Disposable cylindrical clips designed to provide hemostasis to the scalp wound edge.  Part Number: 201037",
      "product_quantity": "906 units",
      "reason_for_recall": "Packaging-pouches with wrinkles across the Tyvek side may compromise the integrity of the pouch's sterile barrier",
      "recall_initiation_date": "20210601",
      "center_classification_date": "20210806",
      "termination_date": "20230127",
      "report_date": "20210818",
      "code_info": "Lot Numbers: W2102045, W2102046, W2102048 and W2102049    GTIN# 10381780511625"
    }
  ]
}