  "results": [
    {
      "status": "Terminated",
      "city": "Bridgend, South Wales",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62745",
      "recalling_firm": "Biomet U.K., Ltd.",
      "address_1": "Waterton Industrial Estate",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA-including the states of IN, SD, and TX and the countries of Austria, China, Japan, Netherlands, Sweden, Turkey, and UK.",
      "recall_number": "Z-2224-2012",
      "product_description": "Oxford Gap Gauge SML 3/4mm.    The instrument is used for an intermediate check of the flexion and extension gaps when preparing the femur to accept the femoral component.",
      "product_quantity": "119 units",
      "reason_for_recall": "A drafting error during an unrelated design change has led to 4 misleading dimensions on the drawing. The CAD model  remained the same thickness but the problem dimensions were accidentally struck from an incorrect point result in an  increase of 1.0mm in the value of each dimension",
      "recall_initiation_date": "20120727",
      "center_classification_date": "20120817",
      "termination_date": "20130215",
      "report_date": "20120829",
      "code_info": "Catalog Number: 32-422801 and Lot Number Identification: ZB111101, ZBll1201, ZB120101, ZB120501, ZB111102, and ZB120201."
    }
  ]
}