      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62641",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Road",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide Distribution - USA including  IN and OH, also, the countries of Australia, Belgium, China, Israel, and Sweden",
      "recall_number": "Z-2223-2012",
      "product_description": "The Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.",
      "product_quantity": "8",
      "reason_for_recall": "1. The current system design of software version 4.0.0xx379 does not prevent the system from scanning if the dose level exceeds a set upper dose limit for CT Brain Perfusion scanning, which previous versions did.  However, the current version does have the Dose Check (NEMA XR-25) feature, which provides dose notifications and alerts, but it is possible for the user to disable Dose Check after ente",
      "recall_initiation_date": "20120718",
      "center_classification_date": "20120817",
      "termination_date": "20130222",
      "report_date": "20120829",
      "code_info": "System Code # 728231, Serial #s:  91003, 95157, 9654, 95130, and 95504    System Code # 728326, Serial #s:  300010, 300030, and 300018"
    }
  ]
}