  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Fairfield",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92527",
      "recalling_firm": "Datascope Corp.",
      "address_1": "15 Law Dr Unit 1",
      "address_2": "N/A",
      "postal_code": "07004-3206",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Domestic distribution nationwide. International distribution worldwide.",
      "recall_number": "Z-2221-2023",
      "product_description": "SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta    Product Description                                                                                            Part Number  SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES\tD684-00-0568-01  SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK)\tD684-00-0568-03  SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE \tD684-00-0568-05  SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY)\tD684-00-0568-09  SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY\tD684-00-0568-01U  PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB  D884-00-0019-22",
      "product_quantity": "321,609 total kits",
      "reason_for_recall": "During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath.  Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.",
      "recall_initiation_date": "20230623",
      "center_classification_date": "20230721",
      "report_date": "20230802",
      "code_info": "Product Description                                                    UDI SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES 10607567108063 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) 10607567109428 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE  10607567109442 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) 10607567113166 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY 10607567109565 PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB   10607567108612"
    }
  ]
}