  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Bartlett",
      "address_1": "3150 Stage Port Drive",
      "reason_for_recall": "When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.",
      "address_2": "Suite 110",
      "product_quantity": "110 units",
      "code_info": "DS50-001-XL-T, Serial Numbers:  24887, 25084, 25085    DS50-002-XL-T, Serial Numbers:  24185, 25219, 25470, 24073, 24187, 24480, 24888, 25089, 24889, 24184, 24813, 24814, 24866, 24867, 24890, 25088, 25220, 24482, 24186, 25090, 24626, 24074, 24189, 25218, 25468, 25469    DS50-003-XL-T, Serial Numbers:  24437, 24260, 24639, 25472, 24257, 24291, 24630, 24640, 24868, 25093, 25478, 24180, 24182, 24259, 24323, 24439, 25222, 24179, 24290, 24293, 24633, 24637, 25091, 25223, 24627, 24077, 24634, 24438, 25476, 25221, 24292, 24263, 24322, 24435, 24628, 24642, 25094, 25479, 24177, 24178, 24641, 25481, 24261, 24076, 24871, 25224, 24078, 24079, 24080, 24175, 24176, 24183, 24256, 24258, 24262, 24264, 24265, 24289, 24321, 24436, 24629, 24631, 24635, 24636, 24638, 24869, 24891, 24892, 24893, 24894, 24896, 24897, 25092, 25095, 25225, 25477    DS50-102-XL-T, Serial Numbers:  24190, 24481, 25467    DS50-103-XL-T, Serial Numbers:  24870, 24898",
      "center_classification_date": "20170602",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany,  Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.",
      "state": "TN",
      "product_description": "Large Stealth Chamber, Catalog Numbers:  DS50-001-XL-T,  DS50-002-XL-T,  DS50-003-XL-T,  DS50-102-XL-T, DS50-103-XL-T",
      "report_date": "20170614",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Iba Dosimetry",
      "recall_number": "Z-2220-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77301",
      "termination_date": "20180611",
      "recall_initiation_date": "20170426",
      "postal_code": "38133",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}