  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Madison",
      "address_1": "441 Charmany Dr",
      "reason_for_recall": "Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.",
      "address_2": "",
      "product_quantity": "15",
      "code_info": "Equipment ID / Serial Number:    MIX-0051 / MX00002051,    MIX-0052 / MX00002032,  MIX-0053 / MX00002028,  MIX-0054 / MX00002052,  MIX-0055 / MX00002043,  MIX-0056 / MX00002040,  MIX-0057 / MX00002050,  MIX-0058 / MX00002036,  MIX-0059 / MX00002027,  MIX-0060 / MX00002015,  MIX-0061 / MX00002016,  MIX-0063 / MX00002039,  MIX-0064 / MX00002033,  MIX-0065 / MX00002053,  MIX-0066 / MX00002014,",
      "center_classification_date": "20170601",
      "distribution_pattern": "US to WI only",
      "state": "WI",
      "product_description": "Exact Sciences Sample Mixer 2.  For In Vitro Diagnostic Use.  The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests",
      "report_date": "20170607",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Exact Sciences Corporation",
      "recall_number": "Z-2219-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77260",
      "termination_date": "20170811",
      "more_code_info": "",
      "recall_initiation_date": "20170428",
      "postal_code": "53719-1234",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}