  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Branchburg",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74094",
      "recalling_firm": "Roche Molecular Systems, Inc.",
      "address_1": "1080 US Highway 202 S",
      "address_2": "N/A",
      "postal_code": "08876-3733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Distributed in NC and WA.",
      "recall_number": "Z-2216-2016",
      "product_description": "cobas¿ Liat¿ Analyzer",
      "product_quantity": "2",
      "reason_for_recall": "Device was released with default pcal/gain settings for  photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.",
      "recall_initiation_date": "20160509",
      "center_classification_date": "20160715",
      "termination_date": "20210324",
      "report_date": "20160727",
      "code_info": "US-IVD: M1-E-00383 & M1-E-00434",
      "more_code_info": ""
    }
  ]
}