  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Tucson",
      "address_1": "250 S Williams Blvd",
      "reason_for_recall": "In Online Entry (OEx) there are two scenarios where a Test result from one container may file to a Test on another container. In order for this to occur:  1)  Site parameter SIT 1,2,1,26 (3,217) [Restrict resulting of tests to owner container (<Y>/N)]=\u0019Y\u0019.  2) The Tests are ordered on the same accession number and   3) The Test codes are defined to have the same upload code as either replicate or",
      "address_2": "",
      "product_quantity": "103 sites",
      "code_info": "Version 6.3.0 and Later",
      "center_classification_date": "20120816",
      "distribution_pattern": "Nationwide Distribution",
      "state": "AZ",
      "product_description": "Sunquest Laboratory SMART & Sunquest Laboratory SMART Select",
      "report_date": "20120822",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Sunquest Information Systems, Inc.",
      "recall_number": "Z-2216-2012",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "62344",
      "termination_date": "20120904",
      "more_code_info": "",
      "recall_initiation_date": "20110428",
      "postal_code": "85711-4472",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}