  "results": [
    {
      "status": "Terminated",
      "city": "Mount Prospect",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65657",
      "recalling_firm": "Rauland-Borg Corp",
      "address_1": "1802 W Central Rd",
      "address_2": "N/A",
      "postal_code": "60056-2230",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US: (nationwide) including the states of  AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NE, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI and  WV., and the countries of : Australia, Canada, Columbia, Lebanon, Mexico and Saudi Arabia.",
      "recall_number": "Z-2213-2013",
      "product_description": "Responder 5 is a nurse call communication system that connects patients with staff in a hospital.  The nurse call system also connects team members and departments from radiology, physical therapy, transport, environmental services and more.        Responder 5 is used in hospitals and long term patient care facilities to allow for remote communications between patients and patient monitoring devices and healthcare providers",
      "product_quantity": "Product Code 351000: 385 units; Product Codes 352000 and 352020: 14,808 units (total)",
      "reason_for_recall": "A counterfeited electronic component was identified in the products affected. The component, a data transceiver (transmitter/receiver), is associated with the intercommunication of these devices in the Responder 5 system across the L-Net data path.   Failure of a data transceiver results in the loss of nurse call functionality in the associated room, and may disrupt communications on the L-Net run",
      "recall_initiation_date": "20130510",
      "center_classification_date": "20130917",
      "termination_date": "20150506",
      "report_date": "20130925",
      "code_info": "1) Product Code 351000 (Branch Regional Controller);   2) Product Code 352000 (Corridor Light);   3) Product Code 352020 (Domeless Controller)",
      "more_code_info": ""
    }
  ]
}