{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Ratisbon",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92534",
      "recalling_firm": "UFSK - International OSYS Gmbh",
      "address_1": "Kirchhoffstr. 1",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-2212-2023",
      "product_description": "UFSK Treatment chair 500 XLE- Treatment chair for positioning and support of the patient during diagnostic and therapeutic treatments of temporary to short-term duration, especially in ophthalmology  Art.no. / REF no.: 15000004  Art.no. / REF no.: 15000006",
      "product_quantity": "239 chairs US",
      "reason_for_recall": "Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position",
      "recall_initiation_date": "20230531",
      "center_classification_date": "20230720",
      "report_date": "20230726",
      "code_info": "Art.no. / REF no.: 15000004 Serial number range: 2000- 2999 Art.no. / REF no.: 15000006 Serial number range: 3000-4506",
      "more_code_info": ""
    }
  ]
}