  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68651",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The product was distributed domestically and to internationally. Domestically to CA, PA, IA, CO, NY, MD, NC,AL, AZ, IL, GA, VA.  Netherlands, Argentina,",
      "recall_number": "Z-2212-2014",
      "product_description": "Device Brand Name: Distal Radius Fracture Repair System  Device Classification Name: Plate, Fixation, Bone  Device Common Name: Drill Guide DNP 3.3mm  Labeling: DRILL GUIDE DNP 3.3 MM NON-STERILE LOT NB0311 QTY 1 DePuy P.O Box 988 Warsaw IN 46581",
      "product_quantity": "37",
      "reason_for_recall": "The firm received complaints indicating that the drill guide becomes stuck in the jig. The drill guide can be placed in the jig about halfway before it becomes stuck.",
      "recall_initiation_date": "20140623",
      "center_classification_date": "20140813",
      "termination_date": "20150714",
      "report_date": "20140820",
      "code_info": "Catalog Number:  DNPDG, Lot Number:  NB0311 and NB0611"
    }
  ]
}