{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Kansas City",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68763",
      "recalling_firm": "Medical Positioning Inc",
      "address_1": "1717 Washington St",
      "address_2": "N/A",
      "postal_code": "64108-1121",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.",
      "recall_number": "Z-2210-2014",
      "product_description": "Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217.      Product Usage:  Used to conduct syncope examinations.",
      "product_quantity": "17 beds",
      "reason_for_recall": "Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal.  The product was manufactured with a component that can cause the device to malfunction.",
      "recall_initiation_date": "20140710",
      "center_classification_date": "20140813",
      "termination_date": "20170103",
      "report_date": "20140820",
      "code_info": "Serial number--802722, 802368, 803147, 804007, 804132, 803195, 802286, 802303, 802782, 803112, 803200, 803274, 803455, 803468, 803883, 803961 and 803960."
    }
  ]
}