  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Branchburg",
      "address_1": "1080 Us Highway 202 S",
      "reason_for_recall": "Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips.  As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run.  No erroneous results would be produced if the issue occurs, but samples in proce",
      "address_2": "",
      "product_quantity": "2,132 boxes",
      "code_info": "Lot Numbers: 11049503, 11053503, 11054503, 11525503 and 11526503.",
      "center_classification_date": "20120815",
      "distribution_pattern": "Nationwide Distribution (USA).",
      "state": "NJ",
      "product_description": "COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.",
      "report_date": "20120822",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Molecular Systems, Inc.",
      "recall_number": "Z-2209-2012",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "62586",
      "termination_date": "20130716",
      "more_code_info": "",
      "recall_initiation_date": "20120201",
      "postal_code": "08876-3733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}