  "results": [
    {
      "status": "Terminated",
      "city": "Kansas City",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68763",
      "recalling_firm": "Medical Positioning Inc",
      "address_1": "1717 Washington St",
      "address_2": "N/A",
      "postal_code": "64108-1121",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.",
      "recall_number": "Z-2208-2014",
      "product_description": "Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017.     Product Usage:  Used to conduct syncope examinations.",
      "product_quantity": "20 beds",
      "reason_for_recall": "Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal.  The product was manufactured with a component that can cause the device to malfunction.",
      "recall_initiation_date": "20140710",
      "center_classification_date": "20140813",
      "termination_date": "20170103",
      "report_date": "20140820",
      "code_info": "Serial number--802971, 803106, 803356, 803588, 803616, 803854, 802781, 802783, 802780, 802784, 802421, 802861, 802939, 802949, 802924, 803867, 803926, 802809 and 804118."
    }
  ]
}