  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Road",
      "reason_for_recall": "Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.",
      "address_2": "",
      "product_quantity": "12 units",
      "code_info": "Serial Numbers:  SN07000141  SN07000109  SN5000162  SN5000421  SN07000667  200148  200115  200081  300114",
      "center_classification_date": "20120815",
      "distribution_pattern": "Worldwide Distribution - USA including IN, MA, MI, MO, NC, NY PA, SC, and WI and to the countries of  Canada, HongKong, New Zealand.",
      "state": "MA",
      "product_description": "Philips Easy Diagnost Classic-Stationary x-ray system  Catalog Numbers: 70642, 70643, 706030, 706031, 706032    Stationary x-ray system, Universal RF system for general use.",
      "report_date": "20120822",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Healthcare Inc.",
      "recall_number": "Z-2208-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62568",
      "termination_date": "20170123",
      "more_code_info": "",
      "recall_initiation_date": "20120711",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}