  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92412",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution. There was government distribution but no military or foreign distribution.",
      "recall_number": "Z-2206-2023",
      "product_description": "Medline O.R. Scissors, S/B, Sterile, 5.5, Model #DYND04000",
      "product_quantity": "189,221 scissors",
      "reason_for_recall": "The scissors were manufactured without a tip protector resulting in the scissors breaking through the semi-rigid plastic tray compromising sterility.",
      "recall_initiation_date": "20230512",
      "center_classification_date": "20230720",
      "report_date": "20230726",
      "code_info": "Lot numbers and UDI numbers:  23BBI031, 01100801968242291023BBI031; 23BBK353, 01100801968242291023BBK353; 23BBP830, 01100801968242291023BBP830; 23BBQ676, 01100801968242291023BBQ676; 23BBR546, 01100801968242291023BBR546; 23CBH009, 01100801968242291023CBH009; 23CBL930, 01100801968242291023CBL930; 23CBP165, 01100801968242291023CBP165; 23BBS369, 01100801968242291023BBS369; 23BBT718, 01100801968242291023BBT718; 23CBV064, 01100801968242291023CBV064; 23CBW834, 01100801968242291023CBW834; 23CBV222, 01100801968242291023CBV222; 23CBW020, 01100801968242291023CBW020; and 23CBQ190, 01100801968242291023CBQ190.",
      "more_code_info": ""
    }
  ]
}